Rapid Determination of Glucosamine Content by Highperformance Liquid Chromatography

XIE Zhi-gang,LI Bo,CHEN Shu-jun,PENG Xiu-lan,WANG Ying,YIN Hong-ping,XI Tao
DOI: https://doi.org/10.3969/j.issn.1004-4957.2012.05.012
2012-01-01
Abstract:A high-performance liquid chromatography(HPLC) method for the determination of glucosamine in glucosamine sulfate potassium chloride capsules was developed.The separation of analyte was performed on a Hedera NH2 column(250 mm ×4.6 mm,5 μm) with acetonitrile and phosphate buffer(65 ∶ 35) as mobile phase,at a flow rate of 1.0 mL/min.The variable wavelength detector(VWD) was set at 195 nm.In the above chromatographic conditions,the calibration curve was linear over glucosamine concentration in the range of 1.6-2.4 g/L.The limit of detection(LOD) of glucosamine was 7.81 mg/L.The RSD(n=6) for the accuracy was 0.53% and the average recovery was 100%with RSD of 1.1%.This method is simple,accurate and with no need of derivation,and is suitable for the determination of glucosamine.