Determination of Maackiain in Sophora japonica by HPLC

Chengli DAN,Yanyan ZHANG,Xiangpei WANG,Hailang LIAO,Shaoji LIU
DOI: https://doi.org/10.3969/j.issn.1001-3601.2016.12.028
2016-01-01
Abstract:In order to determine the content of maackiain in S.japonica,a new high performance liquid chromatography (HPLC) method was established,taking home-made maackiain in laboratory as the reference sample.Diamonsil C18 column (250 mm× 4.6 mm,5μm) was used as stationary phase,ACN-0.1% HAC (40 ∶ 60) with a gradient elution was used as the mobile phase.The column temperature was 25C,the flow rate was 1.0 mL/min,the UV detective wavelength was 310 nm,the injection volume was 20 μL,and the sample was quantified with the external standard method (ESTD).Through the quantified test of six different regions samples of maackiain under the above chromatography conditions.Under these conditions a good linear relationship was obtained with peak area when the sample concentration was 1.328~39.8 ig/mL(R2 =0.999 6),and the average recovery was 102.37% with RSD of 1.59%.In a word,HPLC method established which was used to test the content of maackiain in S.japonica have a series of advantages,such as convenient operations,accuracy,high repeatability and good stability.The content of maackiain in S.japonica sampled in Nanming District reached the highest,100.56 μg/g,which were of little difference in S.japonica in other producing areas,ranging from 57.56 to 79.9 μg/g.
What problem does this paper attempt to address?