Determination of three steroidal saponins from Ophiopogon japonicus (Liliaceae) via high-performance liquid chromatography with mass spectrometry.

Yongyi Wang,Jinzhong Xu,Haibin Qu
DOI: https://doi.org/10.1080/14786419.2012.656109
2013-01-01
Natural Product Research
Abstract:A simple and accurate analytical method was developed for simultaneous quantification of three steroidal saponins in the roots of Ophiopogon japonicus via high-performance liquid chromatography (HPLC) with mass spectrometry (MS) in this study. Separation was performed on a Tigerkin C18 column and detection was performed by mass spectrometry. A mobile phase consisted of 0.02% formic acid in water (v/v) and 0.02% formic acid in acetonitrile (v/v) was used with a flow rate of 0.5?mL?min-1. The quantitative HPLCMS method was validated for linearity, precision, repeatability, stability, recovery, limits of detection and quantification. This developed method provides good linearity (r?>0.9993), intra- and inter-day precisions (RSD <4.18%), repeatability (RSD <5.05%), stability (RSD <2.08%) and recovery (93.82102.84%) for three steroidal saponins. It could be considered as a suitable quality control method for O. japonicus.
What problem does this paper attempt to address?