Precision tilting platform used for static calibration of accelerometers

Xu Guanhua,He Wen,Fu Jianzhong,Xu Yuetong,Lin Zhiwei
DOI: https://doi.org/10.19650/j.cnki.cjsi.2016.11.006
2016-01-01
Abstract:A precision tilting platform used for static calibration of accelerometers was presented.Linear motor was utilized and its linear displacement was converted to the angular displacement of the platform.Simplified mechanical model,as well as measurement and control system for the platform were introduced,and the factors that caused positioning error were analyzed.Static error is caused mainly by temperature variation and the gravity.Manufacturing and assembling error affects the positioning accuracy through theoretical transformation of the linear and angular displacement.Dynamic error is mainly influenced by run out bearing in which parallelism offset of the linear guide rail can be disregarded.Main factors which caused the system error were obtained through experiments,then the compensation method was presented.Results indicate that the dimensional error caused by assembling is the main source of the positioning error.After compensation in software,the absolute error is reduced to less than ±2.0″in the range of -3 600″to +3 600″.
What problem does this paper attempt to address?