Non-contact blood pressure monitoring system based on radio frequency signal

Zeyi TIAN,Xinzhi SHAN,Xiumin GAO
DOI: https://doi.org/10.3969/j.issn.1005-5630.2020.03.012
2020-01-01
Abstract:We proposed a non-contact blood pressure detection method based on radio frequency signals, designed a blood pressure detection system, and conducted data testing and analysis. We mainly foucused on the inconvenience of the traditional cuff sphygmomanometer for the special population and such disadvantages as easily disturbed optical signal, inaccurate measurement, and contacting detection in the existing mainstream methods for detecting blood pressure. The system transmits the radio frequency signal to the point to be detected on the body, and then collects the reflected signal. The waveform of the pulse wave transit can be obtained by the Doppler effect principle, and the pulse wave transit time is calculated and used in the blood pressure calculation model to obtain the blood pressure value, thus achieving non-contact continuous detection and monitoring. The experimental results show that the non-contact continuous detection and monitoring of blood pressure can be realized by the proposed device, which provides a feasible method for non-contact measurement and real-time monitoring of blood pressure by portable and wearable equipment.
What problem does this paper attempt to address?