A Research on Non-Invasive Blood Pressure Measurement System Based on Multi-Sensors

Yun GAO,CongCcong ZHOU,Jian TIAN,Chunlong TU,Xuesong? YE
DOI: https://doi.org/10.3969/j.issn.1004-1699.2015.05.025
2015-01-01
Abstract:Since the blood pressure measurement is susceptible to the arm movement in the process of measurement, this paper implements a noninvasive blood pressure system with an anti-noise algorithm. The system based on the conventional noninvasive blood pressure measurement,collect the real-time information of the arm movement during the measurement by the use of a MEME acceleration transducer,which is used as the reference input of a self-adap-tive filter to eliminate the motion artifact by using the self-adaptive filtering algorithm. The experimental results show that the system is resistant to the partial arm motion artifact in the process of measuring blood pressure measure-ment. The method can be used by the other wearable blood pressure measurement system and improve its anti-inter-ference ability.
What problem does this paper attempt to address?