Characterization and Reduction of X-Gradient Induced Eddy Currents in A Ndfeb Magnetic Resonance Imaging Magnet-3d Finite Element Method-Based Numerical Studies
Xia Li,Ling Xia,Feng Liu,Stuart Crozier,Dexin Xie
DOI: https://doi.org/10.1002/cmr.b.20189
2011-01-01
Abstract:In permanent magnetic resonance imaging (MRI) systems, pulsed gradient fields induce strong eddy currents in the conducting structures of the magnet. The gradient field for image encoding is perturbed by these eddy currents, thus leading to MR image distortions. The conventional approach for the reduction of the eddy current effects is to employ passive shielding (an eddy current board). However, the analysis of the passive shielding structure is typically nonoptimal. In this article, a comprehensive 3D finite element method (FEM) model for the characterization of eddy currents in the ferromagnetic materials of the magnet is developed, which is helpful for the evaluation of the shielding effects and the improvement of passive shielding designs. During the numerical analyses, the eddy current source was introduced through X-gradient coils, nonlinearity, and the anisotropy of the laminated silicon steel structure was considered, with the selection of a penalty factor in the FEM governing equations being also detailed. Based on the simulations results, suggestions have been made for passive shielding designs. (C) 2011 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 39B: 47-58, 2011