Method of core-beam CT imaging scatter correction based on Monte Carlo simulations

Huipeng MENG,Keqiang WANG,Juan FU,Yuanming FENG
DOI: https://doi.org/10.11889/j.1000-3436.2019.rrj.37.040303
2019-01-01
Journal of Radiation Research and Radiation Processing
Abstract:ABSTRACT Based on the EGSnrc program to modularize the core-beam CT (CBCT) system, this study simulated the imaging process entirely based on the accurate construction of the geometric and beam characteristics of the CBCT system. The scatter correction of CBCT images was realized through Monte Carlo simulations by separating the scatter fraction (mainly Compton and Rayleigh scatterings) that affects CBCT image quality based on the interaction between particles and matter. In the phantom images and real cases, the three-dimensionally reconstructed CBCT image significantly reduced the error of the scatter artifacts of the image after scatter correction. In addition, the CT value in the selected region of interest was compared with the planning CT image. The average value of the CT error value in both scan modes was reduced to less than 10 HU, MRSE values were reduced by 71.4%, and CNR values were increased by 19.8%. The scatter correction method significantly improves anatomical details and can be used for CBCT images.
What problem does this paper attempt to address?