Quantitative analysis of Cr in water by LIBS based on internal calibration method

Hui-qin HU,Lin HUANG,Ming-yin YAO,Xiu-wen HE,Ping YANG,Mu-hua LIU,Jin-long LIN,Jing-yi OUYANG
DOI: https://doi.org/10.3969/j.issn.1001-5078.2015.01.007
2015-01-01
Abstract:In order to improve detection accuracy of heavy metal Cr in water by the laser induced breakdown spectros-copy (LIBS),7 kinds of different concentrations of chromium (Cr)water solution were prepared and enriched by the skin cotton,and then the cotton after drying was detected by LIBS.The experimental parameters were optimized,and 425.43 nm was determined as the Cr feature analysis spectral line.The natural calcium (Ca)element in the skin care cotton was chosen as internal calibration element,its feature spectral line is 422.67 nm.Internal calibration method and calibration method were used for quantitative analysis,and the quantitative analysis model was established for five samples,and then the another two samples were used to verify the accuracy of the model.The obtained linear correlation coefficients of the fitting curve are 0.9961 4 and 0.73655 respectively,and the relative errors of two valida-ted samples are 1 1 .38%,4.84% and 64.66%,1 1 .58%;LIBS detection limit is 5.27μg/mL by internal calibration method,and it is better than detection limit 6.06 μg/mL by filter paper enrichment.Experimental results show that the skin cotton as adsorption matrix combined with internal calibration method can improve the accuracy of LIBS in detecting Cr elements in water,and can reduce the detection limit.