Uncertainty in intense pulse electron beam position monitor

Qin Li,Yunlong Liu,Yi Wang,Xinglin Yang,Hong Li,Tiantao Li,Bing Ma,Jin Li
DOI: https://doi.org/10.11884/HPLPB201729.170123
2017-01-01
High Power Laser and Particle Beams
Abstract:Beam position as a key parameter of intense electron beam linear induction accelerator,is criterion of electron beam tuning and experiment.The resistive ring monitor(RRM) and B-dot monitor are often used to measure the beam current and position simultaneously.According to the measuring principle,the calibration method and characters of monitor,type A and type B standard uncertainties and least square fit uncertainty were analyzed as the uncertainty sources of beam position monitors RRM and B-dot,and their evaluation were decided.The combined standard uncertainty of Dragon-Πresistive ring monitor and B-dot is less than 0.1 mm based on calibrating data.