Micro-extrusion Molding of Ultrasonic-assisted Polypropylene Five-lumen Catheter

Zhihong Fu,Yue Wan,Lei Zhang,Gongzheng Zang
DOI: https://doi.org/10.3969/j.issn.1001-3539.2018.04.011
2018-01-01
Abstract:A viscoelastic extrusion model of medical polypropylene (PP) was established,and Polyflow simulation software was used to simulate the gas-assisted extrusion process of ultrasonic-assisted PP five-lumen catheter (the outer diameter is 3.3 mm) by using each lumen diameter and outer diameter of the catheter as the evaluation standards. The effects of ultrasonic power,extruder head temperature and extrusion rate on the dimensions of the various parts of the extruded product under equal injection pressure and equal traction rates were studied. The simulation results for the influence of extruder head temperature on the product size under dif-ferent ultrasonic powers were confirmed through practical extrusion experiment. Simulation and experimental results show that,for PP five-lumen catheter products,there is a maximum of each part size with the increase of the head temperature under the different ultrasonic powers,and the maximum and the corresponding head temperature all decrease with the increase of ultrasonic power. The simulation results show that with the increase of extrusion rate under different ultrasonic powers,the size of each part of the product increases,and the higher the ultrasonic power,the smaller the size of the product.
What problem does this paper attempt to address?