The Determination of Vitamin B12 in Drinks with Immune Affinity Column-Ultra High Performance Liquid Chromatography

Li ZHOU,Yi-gang YU,Hong XU,Chao-zhi HUANG,Wen-ting GUO,Wei-peng GUO
DOI: https://doi.org/10.3969/j.issn.1006-9577.2018.03.006
2018-01-01
Abstract:Super efficient immune affinity column chromatography was adopted to establish the method for determination of the vitamin B12 in vitamin drink,sample after immune affinity column,purification,with 3 mL methanol elution at a speed of flow rate of 1 drop/See,nitrogen blow dry after dissolved by Ultra High Performance Liquid Chromatography (UPLC) method.Chromatographic column for HSS T3,column temperature 35 ℃,mobile phase gradient elution of acetonitrile-0.1% formic acid solution,flow rate of 0.6 mL/min,the wavelength of 360 nm.Results showed that the concentration of vitamin B12 in the 0.4 ~ 10 μg/L range inside sex was good,the regression equation of y =4 040 x-4 040,the correlation coefficient was 0.997 00,recovery rate was 94.68%,the detection limit was 0.1 μg/L This method has the advantages of simple operation,high sensitivity,accurate,high detection efficiency,and is suitable for the determination of vitamin B12 in drinks with complex composition and low vitamin content.
What problem does this paper attempt to address?