Abstract:Effects of key parameters such as gradient elution, flow rate, detection wavelength on the chromatographic response of target analytes were investigated, and the chromatographic conditions for determination of 6 kinds of phthalic acid esters by high performance liquid chromatography with diode array detector were optimized. The ultimate conditions were: chromatographic column:SupelcosilTM LC(15 cm × 4. 6 mm,0. 5 μm); column temperature 25℃; injection volume 15 μL; mobile phase were acetonitrile and water;gradient eluted program:0~11 min V( acetonitrile):V( water) =50∶50, after 11 min V( acetonitrile):V( water) =50∶50; flow rate 0. 8 mL/min; detection wavelength 225 nm. Results indicate that the method has a good linear relationship with all correlation coefficients over 0. 999 8, instrument detection limits of PAEs from 0. 08 to 0. 12 mg/L, relative standard deviations of retention time and chromatographic peak area range from 0. 02%-0. 60% and 0. 13%-0. 86% respectively, which indicated good method precision. The established method is efficient, accurate and repeatable which can be applied to the determination of soil samples.