Determination of Arsenic in Water by Anodic Stripping Voltammetry

Ying Yi,Rilong Zhu,Yueping Luo,Songyang Liu,Yun Tian,Lei Yu
DOI: https://doi.org/10.3969/j.issn.1673-1212.2015.06.026
2015-01-01
Abstract:An In-situ fast analysis of Arsenic in water was performed by anodic stripping voltammetry .The method proposed has low detection limit of 0.007 mg/L and wide detection window between 0.03 mg/L and 0.14 mg/L.For the national standard product, the analytical error can be controlled within 20% and the recovery rate of standard addition only varies from 94.0% to 126.2% by using simple standard determination method .Meanwhile the analytical error was controlled within 5% and the recov-ery rate of standard addition was changed from 93.5% to 104.8% by using standard addition method.The samples with concen-tration of 0.01, 0.02, 0.05, 0.10 mg/L were tested, the RSD is between 7.0% ~18.7%.When the concentration is above 0.05 mg/L, RSD is less than 10%.It is proved that the method could satisfy the needs of in -situ analysis for emergency state.
What problem does this paper attempt to address?