Study on the Rapid Detection Method of Arsenic in Water

Wei HU,Juan DU,Jun CAO,Xin QIAN
DOI: https://doi.org/10.3969/j.issn.1674-6732.2016.04.008
2016-01-01
Abstract:The study was conducted to evaluate the reliability and practical value of rapid arsenic detection method using both labo-ratory and locale experiments.The study showed that the method detection limit was 2 μg/L.The method had good accuracy, the RE was in the range of 1.00%~10.00%when the concentration of arsenic was over 5 μg/L in water.Though the RE was a little high which was 22.40%, when the concentration of arsenic was 5 μg/L, it was acceptable for the rapid detection of arsenic pollu-tion.The method had a good precision and the RSD was in the range of 5.27%~16.54%, and also had a good reproducibility. The method could reduce the workload of pollution screening on arsenic and meet the needs of ansenic pollution incidents.