Evaluation of the surface alterations and the adhesion strength of human periodontal ligament cells on titanium surface treated by pure titanium and conventional ultrasonic scaler tips

Ting-li WEI,Hou-xuan LI,Juan LIU,Chang-xing CHEN,Zhi-bin DU,Fu-hua YAN
DOI: https://doi.org/10.19538/j.kq.2017.07.007
2017-01-01
Abstract:Objective To compare the effects of pure titanium and conventional steel ultrasonic scaler tips on titanium surface and evaluate the adhesion strength of human periodontal ligament cells (hPDLCs)on the titanium surface treated by the two different tips.Methods Totally 60 smooth titanium discs were used in this research.The discs were randomly divided into five groups,each of which contained 12 discs.Ti10 group:discs were scaled for 10s with pure titanium tip;Ti30 group:discs were scaled for 30s with pure titanium tip;Cv 10 group:discs were scaled for 10s with conventional steel tip;Cv30 group:discs were scaled for 30s with conventional steel tip.Control group:discs were not scaled.After the treatment according to the protocols,the morphology and roughness of the discs were evaluated by scanning electron microscope (SEM)and profilometer separately.In order to test the biocompatibility of the treated titanium discs,hPDLCs were co-cultured with those discs for 12 hours,and the cell toxicity of treated discs and the adhesion strength of hPDLCs on titanium discs was calculated with CCK-8 method.Results SEM results showed that Ti10 group had less morphological changes than Cv10 group and this change was highly consistent with the profilometer results which also showed that the roughness of titanium at Ti10 group was significantly lower than Cv10 group (P < 0.05).However,if the scaling time increased to 30 seconds,there were no statistical difference among groups.There were no significant difference among groups for the adhesion strength of hPDLCs after being treated by different scaler tips and no toxicity was found.Conclusion Pure titanium ultrasonic scaler tips are better than conventional ultrasonic scaler tip with respect to the protection of titanium surface,although there were no significant differences in the promotion of cell adhesion.
What problem does this paper attempt to address?