Ex Vivo Estimation of Temperature Combined with Pyroelectric Effect and Ultrasound Attenuation

Yin-fei ZHENG,Yong-gui QIN,Ji-ye AN,Wen-xin FU
DOI: https://doi.org/10.3785/j.issn.1008-973x.2020.01.022
2020-01-01
Abstract:An ultrasonic temperature measurement method combining pyroelectric sensor and temperature-dependent attenuation coefficient was proposed aiming at the influence of the change of ultrasonic signal direction on ultrasonic temperature measurement caused by inhomogeneous medium. A temperature measurement model consisting of ultrasound, test object and pyroelectric sensor was constructed and demonstrated. A sensor with high sound absorption characteristics was designed and fabricated in order to evaluate its response to the ultrasonic signal. Results showed that the output voltage of the sensor was proportional to the reached ultrasonic power. The power sensitivity was 3.117 mV/W, and the measurement error was less than 2.5%. The accuracy of the temperature measurement method was evaluated by estimating the temperature of the pork. The correlation between the sensor output signal time domain energy and the tissue temperature was 0.975 5. The estimated temperature value of the method was basically the same as that of the thermocouple, and the estimated error was less than 3.95 °C. The experimental results show that the pyroelectric sensor and ultrasound can measure the temperature change of the tissue.
What problem does this paper attempt to address?