Key Design Techniques for PVDF Sensitive Element Used in Dynamic Compression Experiments

Xu ZHANG,Shuang QIN,Shuqi YANG,Wenyang PENG,Feng ZHAO,Jun YU,Bin ZHONG
DOI: https://doi.org/10.11858/gywlxb.20190796
2020-01-01
Abstract:This paper introduces key design techniques for PVDF sensitive element used in dynamic compression experiments. Based on the sensitive element design, electrode design, sensitive element polarization and sealing, PVDF dynamic pressure sensors are well design and produced. We also made dynamic experiments in order to calibrate the PVDF sensor. The results show that A-type uncertainty is less than 10%, which can be used in the dynamic pressure range of 0.3–10.0 GPa. We will extend the application to higher pressure range and consider the temperature effects.
What problem does this paper attempt to address?