Analysis of Impurities in Ganciclovir and Ganciclovir Injection

GAO Hui,XIE Yuan-chao,LI Xiao-yan,SHI Guo-sheng
DOI: https://doi.org/10.16155/j.0254-1793.2015.09.24
2015-01-01
Abstract:Objective: To detect and identify the impurities in ganciclovir and ganciclovir injection and to validate the method for the determination of impurities. Methods: The analysis was performed on an Agilent ZORBAX 300-SCX( 4. 6 mm × 250 mm,5 μm) column and the mobile phase was acetonitrile-0. 05% trifluoroacetic acid( 50∶50). The detection wavelength was 254 nm,and the column temperature was at 40 ℃. The target compounds were analyzed with Exactive TM Q mass spectrometry system with HESI in positive ion mode. The spray voltage was 3. 0k V,the capillary temperature was 350 ℃,the atomizer was 250 ℃,the sheath gas was 0. 28 MPa,the auxiliary gas was 0. 1 MPa,and the S-lens RF voltage was 50 V. Results: Ganciclovir was separated completely from known impurities and degradation products under various conditions with the purity factor greater than 999. 0. The largest impurity and the second largest one were guanine and impurity A. It was confirmed that the two major impurities came from synthetic introduction or synthetic process of the material. Conclusion: The method is accurate,sensitive and can be used for determination of the impurities of ganciclovir and ganciclovir injection.