Near-Infrared Off-Axis Integrated Cavity Output Spectroscopic Gas Sensor for Real-Time, in Situ Atmospheric Methane Monitoring

Kaiyuan Zheng,Chuantao Zheng,Haipeng Zhang,Junhao Li,Zidi Liu,Zhiyong Chang,Yu Zhang,Yiding Wang,Frank K. Tittel
DOI: https://doi.org/10.1109/jsen.2020.3042697
IF: 4.3
2020-01-01
IEEE Sensors Journal
Abstract:A near-infrared methane (CH 4 ) sensor system was demonstrated by using off-axis integrated cavity output spectroscopy (OA-ICOS), which was an effective gas-phase analytical technique and was adopted for fast, real-time and in-situ ambient atmospheric CH 4 monitoring. A 60 cm-long cavity with an effective absorption path length of 2150 m was fabricated with a fast response time of ~ 0.8 s. An Allan deviation analysis yielded a minimum detection limit of 13 parts-per-billion in volume (ppbv) using wavelength modulation spectroscopy (WMS) technique, which corresponds to a noise equivalent absorption sensitivity (NEAS) of 9.7 × 10 -10 cm -1 ·Hz -1/2 . By applying multiple denoising scheme of wavelet transform (WT), Kalman filtering (KF) and empirical mode decomposition (EMD) method, the precision levels of CH 4 measurements were improved by a factor of 4, 5 and 6, respectively. For field application, dual-gas sensing was implemented for simultaneous measurement of CH 4 and water vapor (H 2 O) in a single sensor system. Also, a three-day continuous monitoring of atmospheric CH 4 concentration levels was carried out to verify the validity and reliability of the demonstrated ppbv-level CH 4 sensor system.