Design of Performance Test System and Analysis of Temperature Dependence for Space-Borne Array CCD

Yao Pingping,Xu Sunlong,Tu Bihai,Yu Xinyu,Cui Shanshan,Luo Donggen,Hong Jin
DOI: https://doi.org/10.3788/cjl202047.0910001
2020-01-01
Abstract:In response to the demand for the photoelectric performance evaluation of multi-angle polarization imaging detectors, a universal testing system for photoelectric performance parameters of image sensors is developed herein, and a Peltier effect-based thermoelectric cooler and low-temperature cycler are combined to control the temperature of the detector. Experimental results show that the light source stability of the integrating sphere is 0.036% , the uniformity of illuminance is >= 99. 3% , the cooling rate of the detector refrigeration system is approximately 2.2 degrees C / min, and the temperature control accuracy is better than +/- 0.15 degrees C, and the temperature is - 10-25 degrees C . For the measurement of key performance parameters such as quantum efficiency, photo response non-uniformity, dark current, full-well charge, and optical response nonlinearity within the operating temperature range, the photo response non-uniformity of the multi-angle polarization imaging detector at different wavelengths is better than 3 degrees C, The optical response nonlinearity is better than 1% , the dark current increases by 1.25 times when the operating temperature from 0 degrees C to 9 degrees C , and the quantum efficiency in the near-infrared band changes by 3.58 percentage. The performance evaluation of the detector guarantees the test performance and calibration requirements of the multi-angle polarization imaging detector.