Attenuation Compensation for High-Frequency Acoustic-Resolution Photoacoustic Imaging

Haoran Jin,Siyu Liu,Ruochong Zhang,Zesheng Zheng,Yuanjin Zheng
DOI: https://doi.org/10.1109/iscas45731.2020.9180505
2020-01-01
Abstract:Acoustic-resolution microscopy is an important imaging method in studying biological tissues with deep penetration. Using high-frequency transducer could reduce the size of the acoustic focal spot, thereby improving the resolution of microscopy. However, high-frequency photoacoustic signals usually suffer from acoustic attenuation which weakens their energy and distorts the image. In this paper, an attenuation compensation for high-frequency acoustic-resolution photoacoustic imaging is proposed. This technique upgrades the wavenumber term by considering acoustic attenuation and dispersion during wavefield extrapolation, which is a Fourier-domain image reconstruction. It is able to deal with the space-variant attenuation effect and inherits the high computational efficiency of wavefield extrapolation methods. According to the results of simulations and experiments, attenuation compensation successfully eliminates the dispersion induced reconstruction errors, obviously improves the resolution of the image and clearly presents the edges of targets.
What problem does this paper attempt to address?