Wear Simulation of Ceramic-on-Crosslinked Polyethylene Hip Prostheses: A New Non-Oxide Silicon Nitride versus the Gold Standard Composite Oxide Ceramic Femoral Heads.
Makiko Yorifuji,Saverio Affatato,Toshiyuki Tateiwa,Yasuhito Takahashi,Takaaki Shishido,Elia Marin,Matteo Zanocco,Wenliang Zhu,Giuseppe Pezzotti,Kengo Yamamoto
DOI: https://doi.org/10.3390/ma13132917
IF: 3.4
2020-01-01
Materials
Abstract:The purpose of the present study was to compare the wear behavior of ceramic-on-vitamin-E-diffused crosslinked polyethylene (Vit-E XLPE) hip bearings employing the gold standard oxide ceramic, zirconia (ZrO2)-toughened alumina (Al2O3) (ZTA, BIOLOX(R)delta) and a new non-oxide ceramic, silicon nitride (Si3N4,MC2(R)). In vitro wear test was performed using a 12-station hip joint simulator. The test was carried out by applying the kinematic inputs and outputs as recommended by ISO 14242-1:2012. Vitamin-E-diffused crosslinked polyethylene (Vit-E XLPE) acetabular liners (E1(R)) were coupled with o28-mm ZTA and Si(3)N(4)femoral heads. XLPE liner weight loss over 5 million cycles (Mc) of testing was compared between the two different bearing couples. Surface topography, phase contents, and residual stresses were analyzed by contact profilometer and Raman microspectroscopy. Vit-E XLPE liners coupled with Si(3)N(4)heads produced slightly lower wear rates than identical liners with ZTA heads. The mean wear rates (corrected for fluid absorption) of liners coupled with ZTA and Si(3)N(4)heads were 0.53 +/- 0.24 and 0.49 +/- 0.23 mg/Mc after 5 Mc of simulated gait, respectively. However, after wear testing, the ZTA heads retained a smoother topography and showed fewer surface stresses than the Si(3)N(4)ones. Note that no statistically significant differences were found in the above comparisons. This study suggests that the tribochemically formed soft silica layer on the Si(3)N(4)heads may have reduced friction and slightly lowered the wear of the Vit-E XLPE liners. Considering also that the toughness of Si(3)N(4)is superior to ZTA, the present wear data represent positive news in the future development of long-lasting hip components.