A New Tof System for Target Recoil-Ion Momentum Spectroscopy at Imp

Zhou Jiawen,Liu Shubin,Yin Chunyan,An Qi,Zhao Lei
DOI: https://doi.org/10.13538/j.1001-8042/nst.22.372-378
2011-01-01
Nuclear Science and Techniques
Abstract:A high resolution time measurement system with high data transfer rate was designed for the COLTRIMS (Cold Target Recoil-Ion Momentum Spectroscopy) system in Institute of Modern Physics, Chinese Academy of Sciences. It is used to measure the Time of Flight (TOF) with a high resolution for all 20 channels. Based on the PCI Extensions for Instrumentation (PXI) standard, the readout electronics system consists of one Clock-Trigger fan-out module and four TOF modules. Test results show that the system meets the demands of COLTRIMS, with a time resolution of better than 25 ps and a data transfer rate over 20 MB/s.