Simultaneous and Sensitive Determination of As and Sb in Coffee and Tea with Double-Channel Hg-Afs

Yan Hu,Xiaorong Hu,Hui Liu,Long Huang,Rui Liu
DOI: https://doi.org/10.46770/as.2014.02.006
2014-01-01
Atomic Spectroscopy
Abstract:Coffee and tea are two of the most popular beverages in the world. Thus, it is of great importance to develop simple and efficient analytical methods to determine their chemical composition. In this study, a simple analytical procedure was developed for the sensitive and simultaneous determination of trace arsenic and antimony in coffee and tea samples by hydride generation atomic fluorescence spectrometry (HG-AFS). The conditions for the generation of volatile As and Sb species, such as pre-reduction conditions, carrier gas and shield gas flow rates, and reaction media (KBH4 and HCl), were investigated in detail. The precision for five replicates at the 20 ng mL(-1) As level and the 10 ng mL(-1) Sb level was 1.3% and 2.1% (relative standard deviation, RSD), respectively.Under the optimized experimental conditions, the limits of detection (LOD) were 0.072 ng mL(-1) and 0.078 ng mL(-1), with linear ranges of 0-80 ng mL(-1) (R-2= 0.9992) and 0-40 ng mL-1 (R-2 = 0.9965), for As and Sb, respectively. The proposed method was successfully applied for the simultaneous determination of As and Sb in several coffee and tea samples from a local market, and the spike recoveries were in the 92-104% range. The accuracy of the method was validated by analyzing certified reference materials and the results were in good agreement with the certified values.
What problem does this paper attempt to address?