Simultaneous Determination of Rare Earth Elements in Aqueous Samples by Online Preconcentration System and Inductively Coupled Plasma Mass Spectrometry

Zhu Xuxu,Lin Jianjie,Gao Aiguo,Wang Danhong,Zhu Min
DOI: https://doi.org/10.46770/as.2017.04.001
2017-01-01
Atomic Spectroscopy
Abstract:Rare earth elements (REEs) are useful geological tracers, however, their extremely low concentrations (ng/L or lower) and high salt matrix make their measurement a challenge. In this sutdy, an online preconcentration system, which uses a PA -1 mixed resin with ethylenediaminetriacetic acid and iminodiacetic acid functional groups, coupled with a seaFAST autosampler and inductively coupled plasma mass spectrometry (ICP-MS), is used to determine REEs in aqueous samples. The process of the system is automatic, the concentrated samples are directly injected into the nebulizer of the ICP-MS, while the alkali and alkaline earth elements are washed out. The linearity coefficients of the REEs were in the range of 0.99967 (Ce) to 0.99999 (Eu, Gd, Tm), procedural blank values ranged from 0.002 ng/L (Tm) to 0.426 ng/L (Ce), and the DLs (3o) were found in the range of 0.007 ng/L (Tm) to 0.250 ng/L (Ce). The reproducibilities, obtained from 17 aliquots of 500 ng/L diluted
What problem does this paper attempt to address?