The clinical test of domestic linear accelerator AccStar

孙建国,陈正堂,周一兵,李德志,李光辉,李启明,蒲萍,张豫萍,万久庆
DOI: https://doi.org/10.3760/cma.j.issn.0254-5098.2013.04.020
2013-01-01
Abstract:Objective To clinically validate the technical parameters of domestic linear accelerator AccStar,and to assess its accuracy,efficiency and safety in clinical application based on the relevant national standards.Methods Volunteers who suffered nasopharyngeal carcinoma and lung cancer were selected to accept conformal radiation therapy using AccStar medical linear accelerator.A variety of tests were conducted based on the national standards for clinical trials.The clinical efficacy of the radiation therapy was assessed based on the RECIST standard.Acute radiation response was evaluated based on the RTOG standards,and the mechanical and the geometric properties of the accelerator were tested and corrected based on the international standards of IEC and the national standards.Results Forty-three cases were included,of which there were 21 patients with nasopharyngeal carcinoma,and 22 with non small cell lung cancer.All of them received the clinical treatment and auxiliary examination.The overall disease control rate reached 100% and the patients could tolerate the radiation reactions,most of which were skin and oral mucosa reactions.During the clinical trials,all the technical parameters were tested,including the physical parameters and performance parameters,could meet the standards mentioned above.Conclusions The accuracy,efficiency and safety of AccStar meet the relevant national standards and the clinical requirements,so that it can be applied to clinical setting.