Construction & validation of linear elastic model of intraoperative brain deformation during neurosurgical navigtion

Dong-xiao ZHUANG,Yi-xun LIU,Jin-song WU,Cheng-jun YAO,Zhi-jian SONG,Hai-hua ZHU,Chen-xi ZHANG,Man-ning WANG,Wei WANG,Liang-fu ZHOU
DOI: https://doi.org/10.3321/j.issn:1001-2346.2008.10.024
2008-01-01
Abstract:Objietive To compensate for the intraoperative brain deformation for the purpose of increasing the accuracy of neurosurgical navigation. Methods A linear elastic model of porcine brain based on finite element method ( FEM ) was established. After craniotomy in swine,the deformation of cortical surface was tracked by a 3D laser range scanner (LRS) as a boundary condition. This boundary condition was then applied on the finite element equation to simulate the entire brain deformation. Polymethyl methacrylate (PMMA) beads were implanted into the porcine brain as markers of tissue shift,and the predictive accuracy of the model was validated by the real-time data of brain deformation acquired by MR imaging during operation. Results The predictive error of this model ranged from 0.20 to 1.54mm,( mean 0.97±0.44 mm). The accuracy of calibration ranged from 56.5% to 90.0% (mean 68.0±9.6% ). The predictive accuracy on the displacement of superficial markers was higher than that of deeper markers (70.7 ±9.1% : 65.4±10.8%,P < 0.05 ). Conclusions Model-updated image was proved to be efficient,convenient,and reliable in animal research,therefore,it is an ideal approach to compensate for the brain deformation during neurosurgical navigation.
What problem does this paper attempt to address?