Systemic review of the accuracy of implantable Doppler probe for free flap monitoring

Chen Chen,Yuanmin Du,Jingjing Chen,Wei Lu
DOI: https://doi.org/10.3760/cma.j.issn.1674-2907.2016.34.021
2016-01-01
Abstract:Objective To assess the accuracy of the implantable Doppler probe for free flap monitoring. Methods Researches about the implantable Doppler probe for free flap monitoring were retrieved by keywords retrieval in Cochrane, Medion, IFCC, BIOSIS Previews databases and Chinese national knowledge infrastructure ( CNKI) , Wanfang and Chinese biomedical medicine ( CBM) literature database. Two reviewers independently screened the literature, and the Cochrane specified literature evaluation tools were used for quality assessment of the literature. After extracting the data, the Stata software was used to calculate the comprehensive sensitivity, specificity, likelihood ratio, diagnostic odds ratio, and drawing the forest map and SROC curve. Results A total of 568 literatures were retrieved, 9 were included in the study, and the total sample size is 1 136. Meta analysis showed:the sensitivity of implantable Doppler probe for free flap monitoring was 0.93 (0.67-0.99), the pooled specificity was 0.97 (0.96-0.98), the pooled positive likelihood ratio was 36.1 (23.9--54.6), the negative likelihood ratio was 0.07 (0.01-0.4), and pooled diagnostic odds ratio was 498 (79-3 137), and the area under the SROC curve was 0.98 (0.96-0.99). Conclusions The implanted Doppler probe was used to monitor the sensitivity and specificity of free flap monitoring, and they are all high, especially for the detection of early free flap blood supply, and the improvement of the salvage rate of the necrotic flap.