Effects of Amount of Activation for Lingual Bodily Movement of Maxillary Central Incisor on Orthodontic Force Generated by Thermoplastic Aligners

Li Xiaowei,Ren Chaochao,Wang Zheyao,Zhao Pai,Wang Hongmei,Bai Yuxing
DOI: https://doi.org/10.3760/cma.j.issn.1674-5760.2014.04.005
2014-01-01
Abstract:Objective To measure orthodontic force generated by different amounts of activation of lingual invisible aligners with the use of the Micro-stress sensor and explore the possible influences of activation amounts on the orthodontic force.Methods 1.0 mm thickness thermoplastic materials were selected to manufacture the aligners with 0.2 mm,0.3 mm,0.4 mm,0.5 mm and 0.6 mm activation for lingual bodily maxillary central incisor movement.The orthodontic force delivered by the thermoplastic aligners was measured using micro-stress sensor measurement system for invisible orthodontic technique,and monitored force attenuation within 2 weeks.Results The average orthodontic force generated by the aligners with 0.2 mm,0.3 mm,0.4 mm,0.5 mm and 0.6 mm activation was 8.047 N,9.250 N,10.189 N,11.821 N and 12.247 N relatively.All aligners demonstrated a quick decreasing strain at first 8 hours,then the strain decreased slowly and plateaued from day 4.Conclusions As amount of activation increases,orthodontic force generated by the invisible orthodontic appliances also increases.It is recommended that the amount of activation should not exceed 0.5mm for lingual bodily movement of maxillary central incisor.Each aligner should be worn for close to 24 hours per day in the first 4 days.