Resurfacing full thickness defects of focal cartilage in the knee with titanium alloy implant with ceramic coating: an experimental study in sheep

蓝平衡,郭征,吴智钢,李小康,王玲,王财儒,李勇,张涌泉,石磊
DOI: https://doi.org/10.3760/cma.j.issn.1671-7600.2016.08.012
2016-01-01
Abstract:Objective To investigate the efficacy of resurfacing full thickness defects of focal knee cartilage with a titanium alloy implant with ceramic surface coating in sheep.Methods Twelve male sheep were used in this experiment.Standardized defects of focal knee cartilage (7 mm in diameter and 2 mm in depth) were created in the medial femoral condyle at both knees in 9 sheep which were randomly assigned into 3 groups (n =6).In group A,titanium alloy implants with ceramic surface coating were inserted into the defects.In group B,the defects were treated with titanium alloy implants.In group C,the defects were left untreated.The 6 untreated knee joints in another 3 sheep served as a control group (group D).The animals were sacrificed at 6 months.X-ray examinations were performed immediately after operation and at 6 months.Surface roughness of the implants was measured before implantation and at 6 months.The knee joints were evaluated according to O' Driscoll score.The damage to opposing cartilage was evaluated according to Outerbridge score and modified Mankin score.Micro-CT analysis and histological observation were used to assess stability of the implants and the cartilage surrounding the implants.Results At 6 months,the surface roughness value in group B (0.178 ± O.063) was significantly higher than in group A (0.034 ± 0.006) (P <0.05).Group A had significantly higher O'Driscoll score (4.7 ±0.6) than group B (3.8 ±-0.3) and group C (3.7 ±0.3) (P < 0.05).The Outerbridge score (1.0 ±0.4) and modified Mankin score (2.3 ±0.3) for the medial tibial plateau in group A were significantly lower than those in group B (2.5 ± 0.3 and 7.2 ± 1.0,respectively) and in group C (2.0 ± 0.3 and 5.2 ±0.8) (P < 0.05).There were no significant differences between groups A and D (P > 0.05).Micro-CT scan and histological observation showed that all the implants were mechanically stable and new-born cartilage covered the surface of implants well in groups A and B.Conclusion Titanium alloy implants with ceramic surface coating may favor resurfacing full thickness defects of focal knee cartilage in sheep.
What problem does this paper attempt to address?