Evaluation of Repeatability of Higher-Order Aberrations Measurement, in Pseudophakic Eyes by I-Trace Wavefront Analyzer

徐力,刘奕志,何明光,柳夏玲,曾明兵
DOI: https://doi.org/10.3760/cma.j.issn.1006-4443.2008.07.006
2008-01-01
Abstract:Objective To assess the repeatability in measuring higher-order aberrations (HOAs) in intraocular lens (IOL)eyes by a ray trace wavefront analyzer. Methods HOAs were consecutively examined 3 times across a 5.0 ram aperture diameter in 47 intraoeular lens (IOL) eyes using a ray trace wavefront analyzer. Root-mean-square (RMS) values of total higher-order aberrations (THOA), total coma-like aberrations (TC), total spherical-like aberrations (TSA) and total trefoil-like aberrations (TT) were recorded. The repeatability of measurements was assessed by Pearson correlation analysis between each pair of measures and the intra-class correlation coefficient (ICC) among the 3 measurements. The mirror symmetry of HOAs between each pair of the eyes was studied using Pearson correlation analysis. Results In IOL eyes HOAs measurements, all ICC>.73 (P <0.01 ) and all Pearson correlation coefficients r>.57 (P <0.01 ) between each pair of measurements. There was no significant difference in the mean RMS values of THOA,TSA and TT between right and left eyes (P >0.05). But the mean RMS values of THOA,TT had low correlation between right and left eyes. Conclusions The repeatability of i-Trace wavefi'ont analyzer in measuring HOAs in IOL eyes is better. There have good symmetry in the TSA mean RMS values between fight and left eyes, while the mean RMS values of THOA, TC and TT have bad symmetry between right and left eyes.