An Effective Digital Pulse Processing Method for Pile-Up Pulses in Decay Studies of Short-Lived Nuclei

X. Wang,Z. H. Li,Z. Liu,J. Li,H. Hua,H. Y. Lu,W. Q. Zhang,T. H. Huang,M. D. Sun,J. G. Wang,X. Y. Liu,B. Ding,Z. G. Gan,L. Ma,H. B. Yang,Z. Y. Zhang,L. Yu,J. Jiang,K. L. Wang,Y. S. Wang,M. L. Liu,C. J. Lin,L. J. Sun,N. R. Ma,H. S. Xu,X. H. Zhou,G. Q. Xiao,H. Y. Wu,C. Xu,S. Q. Zhang,X. Q. Li,R. Han,Z. Q. Chen,C. G. Wu,D. W. Luo,Y. Jin,J. Lin,D. X. Jiang,Y. L. Ye,F. S. Zhang
DOI: https://doi.org/10.1016/j.nima.2020.164068
2020-01-01
Abstract:With the charged-particle decay spectroscopy extended to the proton drip-line region, half-lives of nuclei of interest decrease rapidly. Consequently, severe signal piling-up resulting from the implanted ions and their successive decays in position sensitive detectors become more and more common. To extract the spectroscopic information from the pile-up pulses with good accuracy, a digital pulse analysis algorithm is presented in this paper and validated with the alpha decay of Th-219 and the internal conversion electron decays of Ra-210,Ra-211 isomers, all of which have similar to mu s half-lives. The results show that with proper baseline correction, standard pulse construction and plateau-region fitting etc., the spectroscopic information of overlapping signals with time separation down to 80 ns and amplitude of individual pile-up pulse down to 70 keV, can be readily extracted.