Relationships Between Haloes and Objective Visual Quality in Healthy Eyes.

Lu Yao,Ye Xu,Tian Han,Linsong Qi,Jiumei Shi,Zhikang Zou,Xingtao Zhou
DOI: https://doi.org/10.1167/tvst.9.10.13
2020-01-01
Translational Vision Science & Technology
Abstract:Purpose:To determine the normal values and relationships between haloes and objective optical quality in healthy eyes.Methods:In this cross-sectional study, haloes, pupillary responses to light, and objective optical quality were measured with the optical quality analysis system (OQAS) and a vision monitor (MonCv3) in 138 right eyes of 138 healthy young men with mean spherical equivalent of 0.32 ± 0.47 D.Results:The mean disc halo size was 77.17 ± 25.03 arcmin. The mean objective optical quality values were as follows: objective scatter index (OSI), 0.58 ± 0.33; Strehl ratio (SR), 0.21 ± 0.05; modulation transfer function cutoff, 36.27 ± 7.98 cpd; OQAS value (OV)100%, 1.21 ± 0.27; OV20%, 0.91 ± 0.23; and OV9%, 0.59 ± 0.16. Disc halo size correlated independently with OSI (P < 0.001) and minimum pupil size (P = 0.003) by forward stepwise regression analysis (disc halo size = 16.60 + 26.24 × OSI + 11.34 × minimum pupil size; R2 value = 17.7%; F = 14.52; P < 0.001).Conclusions:Reference values for disc halo size and objective optical quality in healthy young subjects were established. Eyes with worse objective vision quality exhibited larger haloes.Translational Relevance:The study provided the knowledge and the relationships of OQAS and halo measurements from a well-defined group of healthy young subjects. Both measurements are useful in clinical practice to help quantify the vision quality and complement each other.
What problem does this paper attempt to address?