Beam Extraction System for CYCIAE-14

Ming Li,Tianjue Zhang,Huaidong Xie,Tao Cui,Weiping Hu,Jiuchang Qin,Zhiguo Yin,Yinlong Lu,Lipeng Wen,Jiansheng Xing
2011-01-01
Abstract:A 14MeV medical cyclotron is under design and construction at CIAE (China Institute of Atomic Energy). H ion will be accelerated in this cyclotron and proton beam will be extracted by carbon strippers in dual opposite direction. Two stripping points are chosen in each extracting direction to extract proton beams to different targets to extend the use of the machine and the stripping points can be selected only by rotating the stripping foil. Two modes have been considered of the extraction system, one is designed to be installed on the wall of the vacuum chamber, the other is designed to be inserted vertically from the sector poles. Final choice will depend on the agility, the simpleness and the results of the experimentation. The angle between the stripper and the beam orbit is optimized to improve the extracted beam quality. Numerical simulation shows the two stripping points, the beam orbit and the beam characteristic at each extraction direction. The optimized azimuth of the stripper is also presented in this paper to show its influence for the beam quality. Based on the concept design, the mechanical design and the correlative experimentation have been done; the results are shown in the paper. PACS: 29.20.HM CYCLOTRON;