Influence of Clutters on Ultrasonic Doppler Blood Flow Measurements

Enyao Shang,Jinchun Hu,Shengwu Du,Chang Tian
DOI: https://doi.org/10.1088/1742-6596/1074/1/012162
2018-01-01
Journal of Physics Conference Series
Abstract:Ultrasonic Doppler velocimetry, a non-invasive, low-cost and high-penetration technique, is one of the most suitable methods for measuring blood flow velocity at present. Since previous studies often use the simplified models that do not contain the clutter, users have to design high-performance filters to remove clutters that are irrelevant to blood flow. This paper focuses on the influence of clutters on ultrasonic Doppler blood flow measurements. We establish a signal model that contains the clutter and analyse the influence of clutters on correlation algorithms. Simulation results show that clutters caused by static tissues lower velocity estimators and auto-correlation algorithm is more sensitive to clutters than cross-correlation algorithm.
What problem does this paper attempt to address?