Study and Fabrication of Dynamic Glucose Sensor Based on Single Needle with Three Electrodes

Liang-ze LIU,Rui-feng YUE
DOI: https://doi.org/10.13873/J.1000-9787(2016)02-0078-03
2016-01-01
Abstract:A new type of implantable glucose sensor based on hollow micro-needle is studied and fabricated used for continuous monitoring on human blood glucose concentration. The hollow micro-needle is constituted by two same slot-type stainless steel needles used as working electrode and auxiliary electrode respectively,which glued together with silicone rubber,a Ag/AgCl filament is placed in the hollow micro-needle as reference electrode,and glucose oxidase is placed on the tip of the micro-needle. The experimental results show that linear range of the sensor is 3~22 mmol/L with glucose concentration,sensitivity is 1. 11μA/mmol/L,response time is 10s,and with good anti-jamming performance.