Measurement error in hydrostatic leveling system due to temperature effect and their reduction method

Jian Xu,Zhi-Feng Tong,Yan-Zhao Xu,Hong-Liang Dai
DOI: https://doi.org/10.1063/5.0177885
IF: 1.6
2024-07-01
Review of Scientific Instruments
Abstract:Temperature is a crucial factor influencing the accuracy of the hydrostatic leveling system (HLS), necessitating a temperature compensation test for HLS. This paper investigates HLS temperature compensation through theoretical correction and experimental verification. The influence of temperature on the accuracy of hydrostatic level products is determined through temperature tests on 34 hydrostatic level products. The optimal temperature compensation formula is derived using the non-linear curve fitting method. The HLS is enhanced with a temperature compensation algorithm and temperature sensor, resulting in a new, high-precision, and high-stability hydrostatic leveling product with temperature compensation. Experimental findings reveal that the stability of the improved hydrostatic leveling product exceeds 30% compared to products on the market and surpasses 70% compared to the original products.
instruments & instrumentation,physics, applied
What problem does this paper attempt to address?