Improved Raman Pulse Sequence for Rotation Measurements with Atom Interferometers

Yu-Jie Tan,Miao-Miao Zhao,Pan-Pan Wang,Ya-Jie Wang,Yao-Yao Xu,Xiao-Chun Duan,Cheng-Gang Shao,Zhong-Kun Hu
DOI: https://doi.org/10.1088/1681-7575/ab7174
2020-01-01
Metrologia
Abstract:Since the double-loop atom interferometer consisting of Raman π2−π−π−π2 four-pulse sequences can eliminate the dependence of the interferometric phase shift on the atomic initial velocity, it is widely used for rotation measurements. However, our research shows that by coupling itself with both the atomic initial velocity and the laser detuning, the short pulse duration can contribute to unknown systematic errors within the current instrument sensitivity. The coupling of the pulse duration with the laser detuning inside the device is verified experimentally and presented in this work. An improved pulse sequence design is developed to eliminate these systematic errors and improve the precision of the rotation measurements.