Comparison of Bracketing Calibration and Classical Calibration Curve Quantification Methods in Establishing a Candidate Reference Measurement Procedure for Human Serum 17Β-Estradiol by Isotope Dilution Liquid Chromatography Tandem Mass Spectrometry

Qiaoxuan Zhang,Zhiliang Cai,Yingyi Liu,Haibiao Lin,Qiqin Wang,Jun Yan,Liqiao Han,Jianbing Wang,Peifeng Ke,Junhua Zhuang,Xianzhang Huang
DOI: https://doi.org/10.1016/j.microc.2019.104270
IF: 5.304
2020-01-01
Microchemical Journal
Abstract:A highly accurate liquid chromatography tandam mass spectrometry (LC-MS/MS) based candidate reference measurement procedure (cRMP) for the determination of human serum 17 beta-estradiol (E-2) was developed. A bracketing calibration method (BCM) based on internal standard (IS) concentrations for each sample used to best match the analyte concentration was compared to the classical calibration curve-based method with a constant IS concentration. Precision was analyzed according to the inter-assay reproducibility (n = 15) and intra-assay repeatability (n = 9). Accuracy was confirmed using certified reference materials (CRMs) and further validated by comparison to the established RMPs. Sixty patient serum samples were collected, and the E-2 levels were measured by the BCM isotope dilution (ID) procedure coupled with the LC-MS/MS method to evaluate three immunoassays commonly used in clinical laboratories in China. Limits of quantification (LoQ) as low as 5 pg/mL (18 pmol/L) were achieved. The isomer of E2 (17 alpha-estradiol) was successfully separated by optimizing the LC method. The results of method validation showed that the intra- and inter-assay imprecisions were 1.84 similar to 2.91% vs 2.08 similar to 9.53%, while the analytical recoveries were 98.73 similar to 100.77% vs 93.72 similar to 102.39% for the BCM and classical calibration curve quantification method, and the BCM demonstrated a higher agreement with CRMs and established RMPs. Bland-Altman plots of the E, results revealed concentration-dependent biases. In conclusion, a precise, facile, and reliable ID-LC-MS/MS method is developed in this study. It is demonstrated that the use of the BCM could achieve higher precision and accuracy, which better meet the needs of RMPs.
What problem does this paper attempt to address?