Design and Characterisation of a Novel Z-Shaped Inductor-Based Wireless Implantable Sensor for Surveillance of Abdominal Aortic Aneurysm Post-Endovascular Repair

Nuno P. Silva,Bilal Amin,Eoghan Dunne,Martin O’Halloran,Adnan Elahi
DOI: https://doi.org/10.1007/s13239-024-00753-y
2024-10-09
Cardiovascular Engineering and Technology
Abstract:An abdominal aortic aneurysm (AAA) is a dilation of the aorta over its normal diameter (> 3 cm). The minimally invasive treatment adopted uses a stent graft to be deployed into the aneurysm by a catheter to flow blood through it. However, this approach demands frequent monitoring using imaging modalities that involve radiation and contrast agents. Moreover, the multiple follow-ups are expensive, time-consuming, and resource-demanding for healthcare systems. This study proposes a novel wireless implantable medical sensor (WIMS) to measure the aneurysm growth after the endovascular repair.
engineering, biomedical,cardiac & cardiovascular systems
What problem does this paper attempt to address?