New Nitinol Endovascular Stent-Graft System for Abdominal Aortic Aneurysm with Finite Element Analysis and Experimental Verification

Xiao-Chen Zhou,Fan Yang,Xiao-Yan Gong,Ming Zhao,Yu-Feng Zheng,Zhi-Li Sun
DOI: https://doi.org/10.1007/s12598-019-01250-1
IF: 6.318
2019-01-01
Rare Metals
Abstract:Abdominal aortic aneurysm (AAA) is one of the most common and catastrophic manifestations of the acute aortic syndrome that can be treated with endovascular aneurysm repair (EVAR) which requires a specially designed stent-graft system. In this work, a self-expanding nickel–titanium (nitinol) stent-graft system is aiming at AAA using finite element analysis (FEA) methods to analyze both fatigue behaviors and radial forces. Based on the systematic analysis of the parametric variations, a final stent-graft system was developed by the selection and arrangement of the individual stent components, targeting an optimal performance for the treatment of AAA. Experimental tests, animal tests and clinical trials were carried out to confirm the results. Both animal trials and clinical trials showed comparable curative effects with Medtronic Endurant stent-graft (SG) systems.
What problem does this paper attempt to address?