A Temperature Gradient Field Compensation Method to Improve the Accuracy of the CO 2 Carbon Isotope Sensor

Heng Piao,Defu Cheng,Chen Chen,Yanzhang Wang,Peng Wang,Xiaodong Pan
DOI: https://doi.org/10.1109/tim.2022.3154819
IF: 5.6
2022-01-01
IEEE Transactions on Instrumentation and Measurement
Abstract:Accurate measurement of carbon isotope abundance is extremely important for geochemical applications. In the seismic zone, measurement error caused by variable gas temperature must be compensated. In this study, a 13CO2 carbon isotope abundance sensor using temperature gradient field compensation (TGFC) method is investigated to improve the accuracy of isotope abundance measurement. The theoretical model of sensor is combined with TGFC method to provide a theoretical method for compensating the errors caused by temperature gradient field. When measuring isotopic abundance using the TGFC method, the reference temperature gradient field of the measured gas is obtained through heat transfer analysis and indirect measurement. The reference temperature gradient field of the measured gas is obtained through heat transfer analysis and indirect measurement for sensor measurement using TGFC method. The error of the sensor in $\delta ^{13}\text{C}$ (CO2) is 0.183‰. Two days of tests are conducted to verify the sensor’s performance at the Fengman seismic station. The results validate the ability of the sensor to measure gas isotope abundances in seismic regions for geochemical analysis.
engineering, electrical & electronic,instruments & instrumentation
What problem does this paper attempt to address?