Design and Simulation of the Miniaturized Ion Funnel

Shuguang ZHANG,Jie QIAN,Zaiyue ZHANG,Guangzhou YUAN,Haiyan WU,Lipeng ZHANG,Yi LI,Xiaoxu LI
DOI: https://doi.org/10.16086/j.cnki.issn1000-0380.201801018
2018-01-01
Abstract:Ion funnel is a new type of ion transmission device which greatly improves ion transmission efficiency due to its effectiveness of ion focusing in high atmospheric pressure region.The classical ion funnelis relatively in larger sizes,aiming at this disadvantage,a miniaturized ion funnel has been designed.Through optimizing the geometric parameters,such as the electrode spacing,the ion transmission efficiency is optimized.The electric field simulation software SIMION and ion trajectory simulation software AXSIM are used to simulate the miniaturized ion funnels with different geometric parameters.The simulation results show that when the ion electrode thickness is 0.5 mm and the electrode spacing is 1.6 mm,the ion transmission efficiency reaches the maximum.The effect of electrical parameters on the performance of miniaturized ion funnelis is further studied,and simulation results show that when radio frequency(RF) is 0.5 MHz,RF amplitude is 100 V,and DC voltage gradient is 9.1 V/cm; high transmission efficiency is closed to 95% for ions with mass to charge ratio greater than 100 Da.The optimized miniaturized ion funnel generally has a cylindrical structure,and its length and radial diameter are both less than 36 mm,which can be used in development of miniaturized mass spectrometer.
What problem does this paper attempt to address?