Simultaneous determination of four aflatoxins in dark tea by MFC-IAC coupled to Liquid chromatography tandem mass spectrometry (LC-MS/MS).

Ziling Ye,Pu Cui,Yong Wang,Hangbin Yan,Sanqing Han,Xu Wang,Yu Zhou
DOI: https://doi.org/10.1021/acs.jafc.9b04933
IF: 6.1
2019-01-01
Journal of Agricultural and Food Chemistry
Abstract:Dry tea matrix contains an abundance of caffeine and polyphenols which are different from the food matrix (e.g., protein, lipid, and carbohydrates), and only a few studies have tried aflatoxins determination with tea samples. Here, a specific, accurate, and sensitive method was developed and validated for the simultaneous determination of aflatoxin B1, B2, G1, and G2 in dark teas. Aflatoxins were extracted by acetonitrile/water, press-filtered, and cleaned by multifunctional purification column (MFC) and immunoaffinity column (IAC) in tandem. The cleaned extract was analyzed by liquid chromatography tandem mass spectrometry. The matrix interference was effectively reduced by MFC-IAC cleaning method. Recoveries at the spiking concentrations of 5-60 mu g/kg ranged from 77.5 to 93%, with relative standard deviations <11.0%. The correlation coefficients of aflatoxins standard were >0.998. The limits of detection were 0.024-0.21 mu g/kg and the limits of quantification were 0.08-0.74 mu g/kg. The intra- and interday accuracy ranged from 74 to 87%, and the intra- and interday precisions ranged from 0.4 to 3.1%. After the method validation, the aflatoxins contaminations in 100 collected dark teas were detected, and the results were compared with those of other methods.
What problem does this paper attempt to address?