Digital Pulse Shape Discrimination for Silicon Detector
Pengjie LI,Zhihuan LI,Zhiqiang CHEN,Hongyi WU,Zhengyang TIAN,Wei JIANG,Jing LI,Jun FENG,Hongliang ZANG,Qiang LIU,Chenyang NIU,Biao YANG,Longchun TAO,Yun ZHANG,Xiaohui SUN,Xiang WANG,Yang LIU,Qite LI,Jianling LOU,Xiangqing LI,Hui HUA,Dongxing JIANG,Yanlin YE
DOI: https://doi.org/10.11804/NuclPhysRev.34.02.177
2017-01-01
Abstract:In this paper Pulse Shape Discrimination(PSD) for silicon detector has been briefly introduced. The emerging digital method successfully applied to detector signal processing makes digital PSD method one of the most promising particle identification methods. Sampling frequency and the number of bits are two key parameters of digital method. For silicon detector signal, adopting 100 Ms/s, 12 bit Digitizer can satisfy the time resolution requirement of PSD method. The identification characteristic and energy threshold of this method have been discussed and compared with both front injection and rear injection cases. Energy threshold with rear injection usually is much lower than that with front injection. For example, around for Neon isotope energy threshold with rear injection is about 100 MeV which is only half of the threshold with front injection, also equivalent to thickness of about 60 μm silicon detector threshold in ?E-E method. At the end the impact of silicon detector's resistivity nonuniformity and channel effect on the identification capacity of PSD method has been discussed in detail.