Optimized Fast Simultaneous Non-contrast Angiography and intraPlaque hemorrhage ( fSNAP ) Imaging with low-resolution reference scan for Carotid Artery
Shuo Chen,Xihai Zhao,Jia Ning,Jinnan Wang,Chun Yuan,Huijun Chen
2013-01-01
Abstract:Optimized Fast Simultaneous Non-contrast Angiography and intraPlaque hemorrhage (fSNAP) Imaging with low-resolution reference scan for Carotid Artery Shuo Chen, Xihai Zhao, Jia Ning, Jinnan Wang, Chun Yuan, and Huijun Chen Center for Biomedical Imaging Research & Department of Biomedical Engineering, School of Medicine, Beijing, China, Department of radiology, University of Washington, Seattle, WA, United States, Philips Research North America, Briarcliff Manor, NY, United States Introduction: Simultaneous Non-contrast Angiography and intraPlaque hemorrhage (SNAP) imaging was recently proposed for carotid artery. Previous study has proven its important role for carotid atherosclerotic plaque characterizatoin. By using phase sensitive reconstruction, the artery lumen in SNAP presents negative signal that can be used for non-contrast MRA, while intraplaqu hemorrhage (IPH) presents strong positive signal to be easily detected. However, a full resolution reference acquisition was used in traditional SNAP to obtain background phase for phase sensitive reconstruction, which doubles the total scan time. Recently, a fast SNAP (fSNAP) sequence was proposed to reduce the total scan time by acquiring a separate low-resolution reference image. However, previous proposed fSNAP sequence must be scanned in two separated scan and additional sophisticated phase correction is needed. Moreover, the acquisition efficiency of previous fSNAP still can be improved. Purpose: In this study, we proposed a optimized fast SNAP (fSNAP) sequence with full-resolution inversion recovery (IR) acquisition and low-resolution reference acquisition acquired interleaved in one scan to further speed up the acquisition and simplify the reconstruction for carotid artery. Methods: Theory: The proposed optimized fSNAP sequence is shown in Fig. 1c. Because the background phase has low frequency property. The optimized fSNAP method used interleaved full-resolution IR scan and low-resolution reference scan after an IR pulse to speed up the acquisition. Moreover, arbitrary k-space filling order was also implemented to provide flexible contrast adjustment scheme and further shortening the scan time. IRTR and TI time were made the same as original SNAP. Compared with original SNAP sequence (Fig.1a), the TI and IRTR is preserved to generate similar contrast with shorter scan time in optimized fSNAP. Compared with previous proposed fSNAP (Fig.1b), the optimized fSNAP has higher acquisition efficiency; more importantly, the phase sensitive reconstruction is much simpler without any background phase difference in reference scan by acquiring it in one sequence. Image Acquisition: Three healthy volunteers and one subject with carotid atherosclerotic plaque were recruited in this study. After informed consent, each subject was scanned using a 3.0T whole-body MR scanner (Achieva TX, Philips Medical System, Best, the Netherlands) with a custom-designed 36-channel head&neck coil (11 channel for carotid region). The proposed fSNAP sequence was scanned with parameters: TI 490ms, TR 10ms, IRTR 2000ms, FA11°/5°, FOV 250×160×32 mm, resolution 0.8×0.8×0.8mm, fatsat, NSA 2, turbo factor for IR scan 160, turbo factor for reference scan 40. As a comparison, the original SNAP sequence was scanned with imaging parameters: TI 490ms, linear profile order, TR 10ms, IRTR 1980ms, FA11°/5°, FOV 250×160×32 mm, resolution 0.8×0.8×0.8mm, fatsat, NSA 2, turbo factor 98. Online Reconstruction was used by zero padding the reference acquisition. Image Analysis: The contrast to noise ratio (CNR) of SNAP and optimized fSNAP is calculated