Design and Kinematic Calibration of a New 4-DOF Spinal Surgery Robot
Wenbo Man,Hongjian Yu,Shuo Feng,Zhijiang Du
DOI: https://doi.org/10.1109/robio58561.2023.10354691
2023-01-01
Abstract:This paper presents an analysis of the degrees of freedom, accuracy, and workspace requirements for arch nail implantation surgery, based on which a new spinal surgery robot with a hybrid 4-DOF mechanism is designed and analyzed, and the kinematic parameters are calibrated using the MDH modeling approach. Our design combines a low-degree-of-freedom mechanism with a gimbaled passive joint structure, resulting in smaller dimensions, lower costs, and better flexibility than conventional robots with 5-DOF or higher. As the robot uses a hybrid mechanism, we first analyzed its degrees of freedom and transformed it into a single open chain to obtain an equivalent mDH parameter table to simplify the kinematic problem, and optimized the kinematic parameters based on the Jacobian matrix. Finally, we introduce a fixed-pose inverse kinematic algorithm to analyze the robot’s workspace and dynamic model. Using the degree of freedom analysis, kinematic analysis, workspace analysis, and kinematic parameter calibration methods, this paper presents an analysis method for a multi-directional hybrid robot system and demonstrates that the 4-DOF robot meets clinical requirements. These analysis methods play an integral role in the mechanism analysis, kinematic analysis, and kinematic parameter calibration of the multi-DOF hybrid robot system.