Multicentral Validation of Automatic UWB Radar Algorithm Compared with Polysomnography in Adolescents

Yong Zhou,Kejing Ying,Enguo Chen,Guangyue Qin,Degui Shu,Jianjun Wang
DOI: https://doi.org/10.1183/13993003.congress-2018.pa2250
IF: 24.3
2018-01-01
European Respiratory Journal
Abstract:Multicentral Validation of automatic UWB Radar Algorithm Compared with Polysomnography in Adolescents To validate the accuracy of ultra wide band(UWB) wireless radar algorithm in the screening diagnosis of sleep apnea-hypopnea syndrome.A total of 291 participants from three sleep centers underwent technician attended polysomnography(PSG) and wireless radar device detection simultaneously. There were 115 participants in training group, and their results were used to develop and optimize the apnea-hypopnea index(AHI) algorithm. Other 176 participants were in validation group.AHI results from PSG were reviewed by physicians, while AHI results from radar device were automatically calculated with an embedding chip.UWB radar based AHI algorithm was successfully developed according to respiratory movement and body motion signals. Significant correlation was detected between radar AHI and PSG AHI (Intraclass correlation coefficient 0.983, P=0.000). ROC curve analysis revealed high sensitivity and specificity in all AHI severity groups. High concordance in participants with different gender, age or PSG AHI were reached. The ultra wideband(UWB) radar might be a portable, convenient and reliable device for sleep apnea-hypopnea syndrome screening.
What problem does this paper attempt to address?