Performance Evaluation of a Rotatory Dual-Head PET System with 90̂increments for Small Animal Imaging

F. Meng,S. Zhu,L. Li,J. Wang,X. Cao,X. Chen,J. Liang
DOI: https://doi.org/10.1088/1748-0221/12/09/p09011
2017-01-01
Journal of Instrumentation
Abstract:A rotatory dual-head positron emission tomography (PET) system with 90 degrees increments has been built up by our lab. In this study, a geometric calibration phantom was designed and then used to calibrate the geometric offset of the system. With the geometric calibration, the artifacts in the reconstructed images were greatly eliminated. Then, we measured the imaging performance including resolution, sensitivity and image quality. The results showed that the full width at half maximum (FWHMs) of the point source were about 1.1mm in three directions. The peak absolute sensitivity in the center of the field of view varied from 5.66% to 3.17% when the time window was fixed to 10 ns and the energy window was changed from 200-800 keV to 350-650 keV. The recovery coefficients ranged from 0.13 with a standard deviation of 17.5% to 0.98 with a standard deviation of 15.76%. For the air-filled and water-filled chamber, the spill-over ratio was 14.48% and 15.38%, respectively. The in vivo mouse experiment was carried out and further demonstrated the potential of our system in small animal studies.